The process for Class D medical device registration under the Hong Kong Medical Device Administrative Control System (MDACS) involves several steps designed to ensure the safety, quality, and effectiveness of high-risk medical devices before they can be marketed or used in Hong Kong. Below is a detailed outline of the steps involved in the registration process for Class D devices:
Step-by-Step Process for Class D Medical Device Registration under MDACS
1. Classify the Device
- Class D devices are considered high-risk and include devices that support or sustain life, are implanted, or present a high risk of harm to the patient if they fail. Ensure that your device is correctly classified as Class D.
- Example devices: pacemakers, implantable defibrillators, heart valves, dialysis machines, ventilators.
2. Appoint a Local Responsible Person (LRP)
- The Local Responsible Person (LRP) is a Hong Kong-registered entity responsible for representing the manufacturer. If the manufacturer is not based in Hong Kong, an LRP is mandatory.
- The LRP’s duties include:
- Submitting the application to the Medical Device Division (MDD).
- Acting as a point of contact between the MDD and the manufacturer.
- Handling post-market activities, such as adverse event reporting and recalls.
3. Prepare Documentation
Compile the following documents to support the application:
Application Form MD-14: The official form required for submitting a Class D medical device for listing under MDACS.
Manufacturer’s Evidence of Conformity:
- This can include certification that the device conforms to international safety and performance standards such as ISO 13485 (Quality Management System standard for medical devices).
Certificate of Free Sale:
- This document demonstrates that the medical device is legally sold or distributed in another market with equivalent regulatory standards, such as the US, EU, or Australia.
Declaration of Conformity:
- A document issued by the manufacturer, declaring that the device complies with relevant standards and regulations, such as IEC 60601 (electrical safety) or other international standards.
Risk Analysis Report:
- A report detailing the risks associated with the device and how those risks are mitigated. This report should align with recognized standards such as ISO 14971 (risk management for medical devices).
Clinical Data:
- Documentation that provides clinical evidence of the safety and performance of the device. This is especially important for Class D devices due to their high-risk nature.
Product Description:
- Detailed descriptions of the device, including its intended use, design, and technical specifications. Include information about the device’s components, accessories, and variations.
Labeling and Instructions for Use (IFU):
- Provide examples of the labels, packaging, and user manuals that will accompany the device. This should include details like the intended use, safety warnings, and usage instructions.
Post-Market Surveillance (PMS) Plan:
- A plan for monitoring the device once it is in the market. This should include procedures for tracking and addressing adverse events, recalls, and complaints.
4. Submit the Application
- Submit the completed MD-14 application form and all required documents to the Medical Device Division (MDD) of the Hong Kong Department of Health.
- The application can be submitted in person, via post, or electronically, depending on the available submission methods.
- Submit to the address:
- Medical Device Division, Department of HealthRoom 3101-3102, 31/F, Hopewell Centre, 183 Queen’s Road East, Wan Chai, Hong Kong.
5. Pay the Application Fee
- Pay the required application fee for processing the registration. The fee for Class D devices is generally higher than that for lower-risk devices. The exact fee is outlined in the MDACS guidelines and is non-refundable.
- Ensure that the fee is paid at the time of submission.
6. Application Review by the MDD
- Once the application is submitted, the Medical Device Division (MDD) will review the documentation to ensure that the device meets all the necessary safety, performance, and regulatory requirements.
- During this phase, the MDD may request additional information, clarifications, or supporting documents, such as additional clinical data, technical specifications, or proof of conformity to standards.
- The review process typically involves evaluating:
- The completeness and accuracy of the documentation.
- Whether the device meets recognized safety standards.
- Risk management and clinical data supporting the device’s safe use.
7. Approval and Listing
- If the MDD is satisfied with the application, they will approve the device for registration. The device will then be listed in the MDACS database, which serves as an official record that the device has been evaluated and meets regulatory standards for use in Hong Kong.
- The listing is typically valid for five years, after which it must be renewed.
8. Post-Market Obligations
- After the device is listed, the manufacturer (via the LRP) must comply with ongoing post-market obligations, including:
- Post-Market Surveillance (PMS): Continuously monitor the device in the market to ensure safety and performance. Collect and analyze data from users, including any adverse events or product complaints.
- Adverse Event Reporting: Promptly report any adverse events, malfunctions, or failures related to the device to the MDD.
- Recalls or Corrective Actions: Implement corrective actions or recalls if the device poses a risk to users after being marketed.
9. Renewal of Registration
- Before the expiration of the five-year registration period, submit a renewal application to keep the device listed on the MDACS database. This will involve updating the MDD with any new information related to the device’s performance, safety, or post-market experience.
Summary of Key Steps:
- Classify the Device as Class D (high-risk).
- Appoint a Local Responsible Person (LRP) to act on your behalf in Hong Kong.
- Prepare Documentation, including the application form (MD-14), certificate of free sale, clinical data, and more.
- Submit the Application to the Medical Device Division (MDD).
- Pay the Application Fee for processing.
- MDD Review: The MDD evaluates the application and requests additional information if needed.
- Approval and Listing: Upon approval, the device is listed in the MDACS database for five years.
- Maintain Compliance with post-market surveillance, adverse event reporting, and recalls.
- Renew Registration after five years to maintain the listing.
By following these steps, you can successfully register a Class D medical device under Hong Kong's MDACS system.